Signals appear to be the same on two different channels

* The stimulator creates a high impedance bridge between two catheter pins. If the stimulator is set up the pace catheter pins 1 and 2, the signal from one can be picked up on pin 2 if pin two is removed because there is a direct path through the stimulator at high impedance. This is in place to passively remove polarization from the pacing electrode.
* There is a short in the catheter connector, catheter, amplifier, amplifier cables or junction box.

Ep Medsystems Termination

System will not copy to disk

* Check to make sure the disk is formatted by EPMedSystems. EPMedSystems does not approve of the end-user formatting optical disks on the EP-Workmate as there is a likely potential for data loss or for formatting the hard drive and disabling the system.

Reinhard Schmidt Ep Med Systems West Berlin, NJ (March 13, 2008)--EP MedSystems, Inc. (NASDAQ: EPMD)

The Company’s net loss was $2.1 million, or $.07 per share, for the fourth quarter of 2007, and $5.5 million, or $.18 per share, for fiscal 2007. After excluding business realignment charges, adjusted net loss in the fourth quarter was $787,000, or $0.03 per share, and $4.2 million, or $0.14 per share, for the full year ended December 31, 2007. See the attached reconciliation of the Company’s non-GAAP adjusted diluted earnings per share to the Company’s GAAP diluted earnings per share.

David Bruce commented, “In 2007 we made significant progress growing revenue to record levels, penetrating the ultrasound catheter market, substantially increasing customer satisfaction with our products and services, and developing new products for 2008. Beginning this month we have started shipping our WorkMate® 4.0 products with ClearWave™ technology delivering a significant improvement in clarity of signals, a top factor considered when customers select an EP Recording System.

Ep Medsystems Termination


Monitor has shadows or waves

* Move the cables away from power sources or other cables.
* The video splitter may be defective. Remove it and see if the problem remains.
* The video cables from the main computer to the monitors may be defective.
* Press de-gauss button on the monitor.
* Make sure that there are no magnets or strong magnetic fields near the monitors.

Reinhard Schmidt Termination Intervals are detected too frequently

* Select analysis in the catheter map Setup screen. With the cursor on the analysis type, select the EDIT button. Adjust the Thd. Floor, Thd. Level or Blanking to higher values to increase the accuracy. Adjust the filters to change the location the tick mark is placed. Set the Window to 10.

Don't miss out, learn more: Reinhard Schmidt Termination , Reinhard Schmidt Termination , Reinhard Schmidt Termination





December 1, 2009 | 7:10 AM Comments  0 comments





Reinhard Schmidt's Profile

Reinhard Schmidt's Friends


Latest Posts
Reinhard Schmidt

Monthly Archive
December 2009

Change Language


Tags Archive
epmedsystemstermination reinhardschmidtepmd

Filter By Type
Travel

Friends
Reinhard Schmidt


105 views
Important Disclaimer